Product Question Answer

User Manual


Features


Digital Display Dental Handpiece Clean Maintenance Oil System Lubricating Lubrication Machine







1.Can be maintained for 2pcs high-speed handpiece and 1pcs low-speed handpiece at the same time.
2.Can be maintained for all brands.
3.With the ability to precisely control the time of oiling.
4.With“AIR”key, after the oiling cleaning is finished, you can get out of the excess oil in the handpiece.
5.The door cover is used to effectively reduce the oil mist diffusion and ensure a clean working environment.
6.Intelligent working mode, no need to set the flling time every time.


Technical Parameters

1.Rated voltage:
A:AC220V+ 22V, 50HZ+ 10HZ
B:AV110V+11V,60HZ+ 10HZ(Optional)
2.Rrated power:40W
3.Air pressure:4KG/ cm²-8kg/cm²
4.Work pressure:2.2KG/ cm²-2.5kg/cm²
5.Tank capacity:300ml
6.Motor parameters:
A:Rated voltage :12V ( DC)
B:Rrated power:6W
C:speed:3000rmp/min
7.Working temperature:5-40℃
8.Net weight: 7.0kg


Product Structure

Digital Display Dental Handpiece Clean Maintenance Oil System Lubricating Lubrication Machine


1.Oil station
2.Time display
3.Digital display control area
4.Time setting
5.Switch
6.Air switch

7.Low speed handpiece interface
8.Exhaust
9.High speed handpiece interface 1
10.High speed handpiece interface 2
11.Oil quantity display
12.Sewage pipe

13.Intake pipe
14.Door cover fixing hole
15.Sewage hole
16.Power switch
17.Fuse removal port
18.Power outlet
19.Oil switch


Digital Display Control

Digital Display Dental Handpiece Clean Maintenance Oil System Lubricating Lubrication Machine

A: Press the button "SET" to set the oil filling time of the correspondinginterface. When the corresponding digital flashes, press "+" or "_" toadjust the oil filling time. The filling time can be arbitrarily selected from0 seconds to 50 seconds.(Recommended setting 20s), if you want todirectly work, just press the “SET”button twice. When the digitaldisplay “00", the interface corresponding to this digital does not work.
B:Press the button "START", the lubricator starts working.
C:Press the button “AlR”, the Lubricator to blow the selected interface.and clear the excess oil in the handpiece (if there is still residual oil, youcan press it once more).


Oil Quantity Display&Oil Station

Digital Display Dental Handpiece Clean Maintenance Oil System Lubricating Lubrication Machine

A: There is a oil quantity display bar in the lower left corner, whichshows the amount of oil in the current oil tank. When the oil quantity islower than the green line, it is necessary to add oil to the oil tank. Theoil quantity cannot exceed the red line.
B:The knob in the upper left corner is the tank lid. When the machine is working, you must make sure the seal is on the knob and tighten the knob.


Packing list

1.handpiece lubricator host (1 PCS)
2.Using Instructions (1 PCS)
3.Sewage pipe (1 PCS)
4.Funnel (1 PCS)
5.Fuse(1 PCS)
6.Seal ring(4 PCS)

oil needed for hand piece cleaning machine
SKU TR-D101 ..... wondering what kind of oil should i use in this machine
Hi sir/madam, Thank you for sharing your question. You should add the lubricant in the Clean Maintenance Oil System for handpiece maintenance.

1 Item

Show per page